HPLC Chlorophyll Processing for the ACT Performance Verification of In Situ Fluorometers.
View/ Open
Average rating
votes
Date
2004Corporate Author
Alliance for Coastal Technologies
Status
PublishedPages
1pp.
Metadata
Show full item recordAbstract
The HPLC method used for chlorophyll a analysis followed that of Zapata et al (2000;
MEPS 195:29-45) and was conducted by the laboratory of Dr. Welschmeyer at Moss
Landing Marine Laboratories (MLML, the West Coast ACT Partner Institution). All
pigment samples from Partner sites were collected quantitatively on Whatman 2.5 cm
GF/F filters, frozen in liquid N2 and shipped by overnight courier in liquid N2 dry
shippers to MLML. Samples were removed from liquid N2, extracted overnight in
quantitative volumes (1.8 – 2.0 mL) of N2-purged 90% acetone (-20 oC) and ground with
a motorized Teflon tissue grinder the following morning (the soak-and-grind method).
Samples were kept on ice until loaded on HPLC autosamplers (three identical HPLC
systems were used during this project). Replicate test samples supplied by Partner sites
showed that extraction by the soak-and-grind method yielded as much as 25% more chl a
than by routine acetone soaking alone. Prior to grinding, all samples and.....
Resource URL
Publisher: http://www.act-us.info/evaluations.phpPublisher
Alliance for Coastal Technologies (ACT)Solomons, MD
Document Language
enEssential Ocean Variables (EOV)
Ocean colourBest Practice Type
Best PracticeStandard Operating Procedure
Citation
Alliance for Coastal Technology (2004) HPLC Chlorophyll Processing for the ACT Performance Verification of In Situ Fluorometers. Solomons, MD, Alliance for Coastal Technologies, 1pp. DOI: http://dx.doi.org/10.25607/OBP-360Collections
The following license files are associated with this item: