dc.date.accessioned | 2019-01-29T19:55:04Z | |
dc.date.available | 2019-01-29T19:55:04Z | |
dc.date.issued | 2004 | |
dc.identifier.citation | Alliance for Coastal Technologies (2004) Performance Verification Statement for the YSI Inc. Rapid Pulse Dissolved Oxygen Sensor. Solomons, MD, Alliance for Coastal Technologies, 19pp. (ACTVS04-04). DOI: http://dx.doi.org/10.25607/OBP-377 | en_US |
dc.identifier.other | UMCES Technical Report Series: TS-457-04-CBL / Ref. No. [UMCES]CBL 04-120 | |
dc.identifier.uri | http://hdl.handle.net/11329/821 | |
dc.identifier.uri | http://dx.doi.org/10.25607/OBP-377 | |
dc.description.abstract | Instrument performance verification is necessary so that effective existing technologies can be
recognized and so that promising new technologies can become available to support coastal science,
resource management, and ocean observing systems. To this end, the NOAA-funded Alliance for Coastal
Technologies (ACT) serves as an unbiased, third party testbed for evaluating coastal sensors and sensor
platforms for use in coastal environments. ACT also serves as a comprehensive data and information
clearinghouse on coastal technologies and a forum for capacity building through workshops on specific
technology topics (for more information visit www.act-us.info).
This document summarizes the procedures used and results of an ACT Evaluation to verify
manufacturer claims regarding the performance of the YSI Rapid Pulse Dissolved Oxygen Sensor
incorporated as part of the 6600 EDS. Detailed protocols, including QA/QC methods, are described in
the Protocols for the ACT Verification of In Situ Dissolved Oxygen Sensors (ACT TV04-01), which can
be downloaded from the ACT website (www.act-us.info/tech_evalvations.php). Appendix 1. is an
interpretation of the Performance Verification results from the manufacturer's point of view. | en_US |
dc.language.iso | en | en_US |
dc.publisher | Alliance for Coastal Technologies (ACT) | en_US |
dc.relation.ispartofseries | ACT VS; 04-04 | |
dc.rights | CC0 1.0 Universal | * |
dc.rights.uri | http://creativecommons.org/publicdomain/zero/1.0/ | * |
dc.title | Performance Verification Statement for the YSI Inc. Rapid Pulse Dissolved Oxygen Sensor. | en_US |
dc.type | Report | en_US |
dc.description.status | Published | en_US |
dc.format.pages | 19pp. | en_US |
dc.contributor.corpauthor | Alliance for Coastal Technologies | en_US |
dc.description.refereed | Refereed | en_US |
dc.publisher.place | Solomons, MD | en_US |
dc.subject.parameterDiscipline | Biogeochemistry | en_US |
dc.description.currentstatus | Current | en_US |
dc.description.eov | Oxygen | en_US |
dc.description.bptype | Best Practice | en_US |
dc.description.bptype | Standard Operating Procedure | en_US |
obps.contact.contactemail | info@act-us.info | |
obps.contact.contactemail | tamburri@umces.edu | |
obps.resourceurl.publisher | http://www.act-us.info/evaluations.php | en_US |